The phenotype-o-mat: A flexible tool for collecting visual phenotypes

We’ve developed an easy-to-assemble apparatus and software for the automated collection of visible biological phenotypes such as growth, macroscopic morphology, motion, reflectance, and fluorescence.
Published onApr 24, 2024
The phenotype-o-mat: A flexible tool for collecting visual phenotypes
·

Purpose

Many projects at Arcadia require us to collect visual phenotypes from Petri dishes or multiwell plates, but the existing tools for this are often cumbersome, expensive, and difficult to adapt to new protocols. We therefore created an easily extensible imaging system by combining a low-cost but high-quality camera, a microcontroller-mediated multi-wavelength illumination system, and a software package for writing data acquisition protocols.

We’re sharing the information necessary to assemble the “phenotype-o-mat,” as well as the software required to operate it and example “protocols” written with that software. This pub may be helpful to anyone interested in collecting visual biological phenotypes.

Share your thoughts!

Watch a video tutorial on making a PubPub account and commenting. Please feel free to add line-by-line comments anywhere within this text, provide overall feedback by commenting in the box at the bottom of the page, or use the URL for this page in a tweet about this work. Please make all feedback public so other readers can benefit from the discussion.

The strategy

As part of the platform effort “Genetics: Decoding evolutionary drivers across biology,” we’ve created a hybrid population of the photosynthetic algae Chlamydomonas by crossing C. reinhardtii and C. smithii. Motivated by some of our previous work [1][2] suggesting analysis of many high-dimensional phenotypes will aid in genotype-phenotype mapping, we’re conducting high-throughput genotype and phenotype analysis on this hybrid population. Accordingly, we sought to create a flexible and inexpensive tool for collecting a broad range of phenotypes (e.g. growth rate, colony morphology, chlorophyll fluorescence) from these strains. The design would need to be easy to program, extensible, and potentially useful for other projects at Arcadia. Our solution, described in this pub and a companion protocol, is a simple imaging device with trans-illumination, incident illumination of multiple wavelengths, an optical filter slider for fluorescence assays, and complementary software intended to make programming assays on the device easy.

The problem

For many studies, collecting biological measurements is a bottleneck. Assays often require expensive specialized equipment and experimenter supervision, and they can be difficult to adapt to new biological questions.

Our solution

We developed an inexpensive and easily extensible tool for the automated collection of image-based biological measurements. The “phenotype-o-mat” (Figure 1) consists of a quantitative camera, white-light transmission illumination, several wavelengths of incident illumination, a filter slider, and software infrastructure for programming the camera and illumination system to conduct assays such as cellular growth curves, reflectance, fluorescence, videos, and time series.

Figure 1

Schematic of the phenotype-o-mat.

Approximate costs of individual parts are noted. Components with an asterisk (*) are 3D-printed.

Why is this useful?

The phenotype-o-mat allows the collection of image data including reflectance in four wavelengths, fluorescence, cell density, colony morphology, and small animal behavior (e.g. nematodes or flies). It can collect data in densely sampled videos (up to ~1,000 frames per second) or in arbitrarily long time-lapse imaging sessions. The camera is very sensitive and provides quantitative data. For people in need of different functionality (e.g., different frame rates, resolutions, pixel sizes, quantum efficiencies, or polarization sensitivities), there are other cameras available from the same manufacturer that are compatible with the same software. For fluorescence assays, we add a filter to the front of the camera. Inexpensive filters for many wavelengths are widely available, allowing easy extension of the system’s functionality.

Alternatives to this system such as plate readers or multi-wavelength imagers are expensive and less flexible. For example, adding a new wavelength of incident illumination to the phenotype-o-mat is as simple as ordering a few more light-emitting diodes. Adding wavelengths to a plate reader or imager could cost thousands of dollars or may simply be impossible.

The resource

The phenotype-o-mat is a basic imaging system that you can build yourself from a mix of 3D-printed and commercially available, inexpensive parts (Figure 2). For detailed, step-by-step instructions on how to assemble the device yourself, check out our accompanying protocol. If you run into any issues, please comment on the protocol or the pub and we’re happy to talk you through it.

This section provides an overview of the components required for the device, a quick summary of how to assemble it, some of the decision-making behind its design, and two example assays that it can perform.

All associated code, including example analysis scripts, is available on GitHub (DOI: 10.5281/zenodo.11050006). Data collected in our example assays, including raw image and video files, are available on Zenodo (DOI: 10.5281/zenodo.11043291).

Figure 2

Assembling the phenotype-o-mat.

Overview of the required components and their assembly. For full details, see our assembly instructions on protocols.io.

Components

To develop our imaging system, we first identified cameras that would be quantitative and flexible. We chose the Blackfly S line of cameras from Teledyne. There are many types of imaging chips available in the Blackfly S format, enabling a researcher to optimize for their individual needs. Some cameras have high frame rates, high quantum efficiency, high resolution, low read noise, or specialized features such as simultaneously imaging different polarizations. Furthermore, they all interface with the same software development kit, and so will all be compatible with the software we’ve developed.

We chose the Arduino line of microcontrollers to interface with the illumination system and other possible future peripherals, like an automated filter wheel or a vibrational stimulator for the samples. The Arduino microcontroller ecosystem is well-established and supported. The controllers themselves are inexpensive, easy to program, and have a wide range of functionality.

With the basic components established, we then decided on a core set of imaging functionalities for the phenotype-o-mat: transillumination, incident illumination of four wavelengths, and the ability to use filters for fluorescence assays.

Assembly

A components list and comprehensive instructions on the assembly of the phenotype-o-mat is available on protocols.io.

Briefly, you begin by 3D-printing the physical components. You can also order these here. Next, you’ll add threaded inserts to the plate-holder base and camera arm with the help of a soldering iron. To assemble the illumination ring, you solder current-modulating resistors to each light-emitting diode, then solder these diodes to the printed circuit board for the illumination ring. Next, solder wires connecting the illumination ring to the microcontroller. You go through a similar process for the transillumination panels. You then assemble the physical apparatus by:

1) fitting the 20 mm × 20 mm aluminum extrusion into the plate-holder base and securing it by adding thumb screws into the threaded inserts, 2) inserting the transillumination LED panels into the base, 3) adding the camera arm to the extrusion and securing it with thumbscrews, 4) attaching the illumination mount and the LED ring and diffuser, and 5) adding the camera mount and the camera itself. After physical assembly, you will install the firmware on the microcontroller and the imaging software.

Example assays and data analysis

In this section, we provide some example assays we’ve conducted with the phenotype-o-mat, providing a sense of the breadth of its utility. In our first example, we quantify the reflectance of four different wavelengths of light by two Chlamydomonas species: C. reinhardtii and C. smithii. In this assay, the species are arrayed in 96-well format on solid media plates. In the second example, we measure the decay in chlorophyll fluorescence in these same two species streaked onto Petri dishes. This fluorescence decay is induced when dark-adapted organisms are exposed to light for an extended period. The decay captures biophysical and biochemical responses that allow the organism to adapt to continuous light exposure [3][4].

You can find analysis scripts that generated the technical components of these figures on GitHub, and the data are available on Zenodo.

Multi-wavelength reflectance and colony size

Spectral reflectance patterns have been used to identify plant and microbe species [5][6]. This has been a powerful method in the remote sensing field, but little used to evaluate the phenotypes of different strains in a laboratory setting. The phenotype-o-mat can evaluate reflectance in four wavelengths of light: 460 nm, 535 nm, 590 nm, and 670 nm.

As part of the platform effort “Genetics: Decoding evolutionary drivers across biology,” we created a hybrid population of the photosynthetic algae Chlamydomonas by crossing C. reinhardtii and C. smithii. We’ve subjected this population to phenotypic and genetic analysis with high-throughput methods and have spent significant effort to characterize the phenotypes of the two parental species [7].

In this assay, we sought to determine if the two parental strains yielded differing spectral reflectance patterns. If we were to find variation, we would then use these patterns as a complex phenotype in our Chlamydomonas hybrid population.

We grew Chlamydomonas colonies on TAP media [8] at room temperature and 24-hour illumination arrayed in 96-well format on a single-well microtiter plate. We positioned this microtiter plate in the plate holder at the base of the phenotype-o-mat. We then covered the imaging system with a dark box and ran a script that conducted the following steps:

  1. Turn on the transillumination light source.

  2. Acquire a single image to allow colony segmentation and turn off the transillumination.

  3. Turn on one of the four incident illumination wavelengths available.

  4. Image the colonies and turn off the incident illumination.

  5. Repeat starting at step three until we’ve imaged colony reflectance at all wavelengths.

The script implementing this experimental regimen is available on GitHub.

Figure 3

Spectral reflectance data we collected with the phenotype-o-mat show that C. reinhardtii reflects more light than C. smithii.

(A) A raw image from the phenotype-o-mat (left) and a processed version in which colony segmentation is indicated by shading (right).

(B) Schematic depicting multi-wavelength incident illumination, the setup used to collect the data in C and D.

(C) Four images (one for each wavelength, left to right) for each of 12 representative individual colonies of each species. Colors below columns indicate wavelength (blue = 460 nm, green = 535 nm, yellow = 590 nm, and red = 670 nm). Horizontal arrows indicate examples with variation in reflectance between colonies and vertical arrows indicate examples with variation across wavelengths.

(D) Mean pixel intensities for both Chlamydomonas species we tested across four different wavelengths of incident light. au = arbitrary units. An asterisk (*) indicates a significant difference (p < 0.01, two-tailed t test).

Analyses of data we collected via this approach for C. reinhardtii and C. smithii are shown in Figure 3. We used transillumination images of single-well microtiter plates where colonies are organized in 96-well format (Figure 3, A, left) to identify and segment individual colonies (Figure 3, A, right).

We then propagated these segmentations across each of the four incident illumination images of these same microtiter plates. Representative individual colonies for each of these four images (illumination wavelengths indicated at the top of each column) for both species are shown in Figure 3, B. In these images, surface reflectance (glare) is shown in black. We noticed positional effects, likely from uneven illumination, in reflectance where individual wells exhibited more or less reflected light across all wavelengths (compare horizontal arrows), which will contribute structured noise to our data. But, we also noted that there was consistent reflectance of specific wavelengths across individual colonies (compare vertical arrows) likely indicating that, for any given species, there is consistent variation of reflectance across wavelengths independent of well position.

While we could statistically control for the types of variation caused by colony position, for this initial work we took an agnostic approach and simply compared the difference in reflectance between C. reinhardtii and C. smithii without controlling for possible confounding variables. This analysis is shown in Figure 3, C. We found that, for all wavelengths, C. reinhardtii (red, left) reflected more light than C. smithii (green, right, p < 0.01, two-tailed t test). Likely, these effects would be further accentuated and resolved if we controlled for colony position to account for the uneven illumination.

Chlorophyll fluorescence decay

In dark-adapted plants and photosynthetic algae, exposure to light induces an increase in fluorescent light emitted by chlorophyll A contained in photosystem II [3][4]. The increase in light emission and the subsequent decay that occurs with sustained light exposure contains characteristic indicators of the photosynthetic process, including the efficiency of photosynthesis [9] as well as the photochemical [10] and non-photochemical [11] mechanisms that protect photosynthetic organisms from over-exposure to light. Furthermore, researchers have used both fluorescence induction and decay to guide genetic screens for genes involved in photosynthesis [12]

We sought to develop an assay for the phenotype-o-mat that would allow us to efficiently assess chlorophyll fluorescence decay such that we could analyze the slower components of fluorescence decay in our Chlamydomonas hybrid population described above. 

For this assay, we positioned a 650 nm longpass filter in the light path, allowing only the fluorescence (mostly from chlorophyll) at wavelengths above 650 nm to access the camera. We placed a single-well microtiter plate containing 96 Chlamydomonas colonies growing on TAP media into the plate holder at the base of the phenotype-o-mat. We then covered the imaging system with a dark box and ran a script that conducted the following steps:

  1. Wait 15 min for the Chlamydomonas to dark-adapt [12].

  2. Turn on the 460 nm incident illumination LED.

  3. Once a minute for 20 min, collect 5 s of video at 260 Hz.

The script implementing this experimental regimen is available on GitHub.

Figure 4

Chlorophyll fluorescence decay data we collected with the phenotype-o-mat show that C. reinhardtii and C. smithii differ in their rate and intensity of decay.

(A) Schematic of the experimental protocol used to collect the data in C.

(B) Segmented images of C. smithii, (top) and C. reinhardtii (bottom). Segmented colonies passing filters are indicated by dots. Clearly separated colonies that our process failed to identify are shown in circles and light reflections incorrectly classified as colonies are shown in triangles.

(C) Average decay trajectories for C. smithii colonies (gray) and C. reinhardtii colonies (salmon). Time points are the average fluorescence intensity of all segmented colonies at each acquired time point. The white box with an asterisk (*) contains time points that differ significantly between the two species (p < 0.01, two-tailed t test).

We then calculated the average emitted light intensity for each colony at each time point (Figure 4). For this experiment, and in contrast to the reflectance data, we imaged C. reinhardtii (Figure 4, A, top) and C. smithii (Figure 4, A, bottom) as individual colonies streaked on TAP media in Petri dishes. To identify colonies, we segmented the first image from a series (Figure 4, A, first column) based on fluorescence intensity and filtered based on aspect ratio and colony size. Colonies passing filtration for each species are indicated in red in Figure 4, A, second column. This process was imperfect — several well-separated individual colonies (examples indicated by circles in Figure 4, B) were not identified by this approach and, in some instances, reflectance from the media was classified as colonies (examples indicated by triangles in Figure 4, B). Nonetheless, using this acquisition and analysis, both C. reinhardtii and C. smithii have a fluorescence trajectory consistent with previous findings, where there is an initial spike in emission followed by a slow decay (Figure 4, B) [3][4]. However, the two species significantly differed in the rate and complexity of this decay (Figure 4, B, significant differences between time points indicated by bar, p < 0.01, two-tailed t test). This result suggests that this approach to quantifying chlorophyll fluorescence decay is useful in categorizing individual species and that it will be interesting to investigate the segregation of this phenotype in our Chlamydomonas diversity population.

All analysis scripts that generated these figures are available on GitHub, and the data are available on Zenodo.

Next steps

The phenotype-o-mat is intended as an extensible system and we plan to continue expanding its functionality as new needs arise. For example, these analyses used a version of the apparatus with a commercially available imaging stand, while our newer version contains a more flexible 3D-printed imaging stand and produces similar data.

While we don’t plan to modify the device in the immediate future, we’ll share updates when they come and we hope you'll try developing assays for your own research. If you have questions for us or thoughts to inspire other readers, please comment on the pub! We’d especially love to hear from you if you try building a phenotype-o-mat and have any feedback.


Share your thoughts!

Watch a video tutorial on making a PubPub account and commenting. Please feel free to add line-by-line comments anywhere within this text, provide overall feedback by commenting in the box at the bottom of the page, or use the URL for this page in a tweet about this work. Please make all feedback public so other readers can benefit from the discussion.


Contributors
(A–Z)
Visualization
Methodology, Validation
Validation
Conceptualization, Formal Analysis, Investigation, Methodology, Software, Visualization, Writing
Validation
Supervision
Connections
1 of 3
Comments
3
?
Sachin Jain:

Congratulations on developing this impressive device—the cost-to-benefit ratio is truly remarkable. It's incredibly inspiring for DIY labs. I work with filamentous fungi, and I’m curious if you’ve tested it with fungi yet. One of the challenges I face with fungi is their tendency to form large colonies that expand radially from the point of inoculation. It would be extremely helpful to track their radial growth over time on agar plates. It is one of the ways to measure radial growth rate between different strains of fungi. Additionally, I’m wondering if the device can be placed in incubators to accommodate organisms that require specific temperatures, either high or low, for optimal growth.
Thank you!

?
David G. Mets:

Hi Sachin,

We have not done any work with fungus yes, however, we have measured radial colony growth on Chlamydomonas so it should work for fungus too! We have also used it to calculate growth curves for liquid culture. It can easily be placed in an incubator, deli-case, biosafety cabinet, etc…

?
Rex Kerr:

Cool little device!

A couple of ideas for simple improvements, if they are ever needed: (1) The transillumination doesn’t look very flat in figure 1, mostly because of the boundary effect. A strip of aluminum foil, polished side in, at the front and back ends of the plate holder might be an easy way to improve this enough to matter. (2) If you need better collimation of the transillumination light, you could add an extra little slot above the light panels for a pair of crossed laptop privacy filters (cut to size—a guillotine-style paper cutter usually works nicely). For samples that scatter but aren’t highly-absorbing, this can increase contrast a lot.

Ben Braverman:

Hi Rex, thank you for your thoughtful suggestions. I assume you are referring to transillumination not looking very flat in Figure 3A? If so, we think that the dimmer regions you see at the edges of the plate may actually be coming from the imaging system itself (vignetting from the lens). We are using a cheaper lens to cut down on costs, but upgrading to a higher quality lens or possibly a flat-field corrected lens might help with this. One problem we do have with the transillumination is non-uniformity at specific points in the illumination field. We are using 3 separate LCD backlights and there are (not surprisingly) sharp discontinuities at the boundaries of the back-light panels. We have not tested a laptop privacy filter for polarizing the transillumination light, but that is a great idea! We haven't yet used the phenotype-o-mat for acquisition of things like nematode motility for example, but it might really help in that case.

?
Jose Aguilar-Rodriguez:

In principle, this system could also work with 384- or even 1536-well formats. Is that right? A particularly intriguing "phenotype" to map with this phenotyping method would be evolvability—the potential of a population to genetically adapt to new selective pressures. This could be quantified by measuring the rate of fitness increase over the course of an evolution experiment, using a proxy for fitness such as colony size. Indeed, mapping the rate of change in any measurable and relevant trait could also provide valuable insights. With this approach, hundreds or even thousands of independent populations could be evolved by, for example, serially pinning them down in a 384- or 1536-spot format on agar plates every 24 hours using a Singer rotor. Continuous imaging with multiple phenotype-o-mat setups would allow real-time monitoring of how fitness or other traits evolve during the experiment, depending on the starting genotype of each population.

An example of how evolvability can be mapped is provided in DOI: 10.7554/eLife.27167. Furthermore, using a panel of yeast strains that allows for QTL mapping at single-nucleotide resolution, as described in DOI: 10.1016/j.cell.2017.12.015, evolvability loci could potentially be mapped at the nucleotide level.

?
David G. Mets:

With regard to the technical question, yes! We have collected a fair amount of data from 384 well plates. One issue with the 1536 well plates is the ratio between the size of the bottom of the well and the well wall. Because the system is not built with a flatfield objective (this was done to keep the cost down), large portions of the wells at the edges of a 1536 plate will likely not be visible. That said, using agar (as you suggest) in these formats should be fine and we have conducted several growth assays using colony size on agar plates under differing environmental conditions.

WIth regard to the issue of mapping evolvability (and also phenotypic plasticity in the face of environmental change), this is a great idea, something we have discussed extensively, and the exact type of high throughput experiment that the phenotype-o-mat was designed to enable.